Perioperative medicine & outcomes

Clinical validation of an intra-abdominal pressure monitoring system by comparison with pneumoperitoneum pressure

J Clin Monit ComputPublished prospective cohort study

Authors' conclusion

The IAP monitoring system underestimated the IAP set by the CO2-insufflator by approximately 1 mmHg. The wide limits of agreement warrant cautious interpretation. Overall, the IAP monitoring system showed reasonable accuracy and precision. The increase of several renal biomarkers during laparoscopy compared to baseline may reflect a direct effect of elevated IAP on the kidney.

Study context

Population
Patients undergoing laparoscopy at the University Medical Center Groningenpatients, 99 measurements performed between March 2025 and January 2026
Participants
29 patients
Exposure
Intra-abdominal pressure (IAP) monitoring system compared to a reference standard, the CO2 insufflator, during laparoscopy
Outcome
Agreement between the IAP monitoring system and the CO2-insufflators (Bland-Altman, accounting for repeated measures)

Main finding

As reported in the abstract

Bias

−1.4 mmHg (95% CI (−1.96 - (−0.82))

Limits of agreement

- 5.96 mmHg and 3.2 mmHg

95% CI −6.8–3.99

Precision

2.3 mmHg

Observational data: associations, not effects.

Further findings

Renal biomarkers (KIM-1, TIMP-2, albumin/Cr ratio) · during laparoscopy compared to baseline

Levels of KIM-1, TIMP-2 and albumin/Cr ratio increased during laparoscopy compared to baseline

Type of surgery

Hemicolectomy
One patient (3%)
Endometriosis surgery
28 patients (97%)

Original publication

Clinical validation of an intra-abdominal pressure monitoring system by comparison with pneumoperitoneum pressure. J Clin Monit Comput. Published 14 October 2026. doi:10.1007/s10877-026-01515-y

Composed by AI from the published abstract and checked against it. How Briefs are made →