Development and validation of a medical latex glove-type real-time cardiopulmonary resuscitation feedback device

ResuscitationPublished Randomised crossover simulation study

An erratum to this article was published (Resuscitation, May 2026 ↗)

Authors' conclusion

Chest compressions with the novel medical latex glove-type CPR feedback device resulted in a greater adherence to the device-displayed guidance on chest compression depth than the conventional device did.

Study context

Population
Simulated 2-min chest compressions on a manikincrossover, 3 conditions
Participants
30 participants
Comparison
No feedback vs CPR feedback device (CFD) vs SFD
Primary endpoint
User adherence to device-displayed guidance (proportion of adequate compression depth 5-6 cm and rate 100-120/min) and accuracy of device-displayed value vs manikin reference data
Methodological detailsrandomized
  • Randomized

Original publication

Development and validation of a medical latex glove-type real-time cardiopulmonary resuscitation feedback device. Resuscitation. Published 9 February 2026. doi:10.1016/j.resuscitation.2026.111001

Composed by AI from the published abstract and checked against it. How Briefs are made →