Evaluation of non-invasive sensors for monitoring core temperature

J Clin Monit ComputPublished Prospective cohort study

An erratum to this article was published (J Clin Monit Comput, Dec 2025 ↗)

Authors' conclusion

Neither the ZHF nor DS non-invasive probes provided sufficient accuracy or precision to guide clinical decisions in the ICU.

Study context

Population
Intensive care unit (ICU) patients undergoing hypothermic temperature control358 observations
Participants
35 patients
Exposure
Zero-heat flux (ZHF) and dual sensor (DS) non-invasive temperature probes vs direct core temperature measurement
Outcome
Bias (accuracy) of non-invasive probes vs direct core temperature (Bland-Altman); precision by Lin's concordance coefficient (LCC)
Methodological detailssingle-centre, method comparison (Bland-Altman)
  • Single-centre
  • Method comparison (Bland-Altman)

Main finding

As reported in the abstract

Mean bias vs invasive temperature

ZHF

+ 0.98

mean · °C

DS

- 2.19

mean · °C

Observational data: associations, not effects.

Further findings

OutcomeZHFDS
Bias and LCC in hypothermic patients (subgroup)+ 0.28 °C- 2.52 °C
LCC0.860.07

Clinical confounders (vasoactive agents, temperature control devices)

Did not consistently affect bias, accuracy, or precision

Original publication

Evaluation of non-invasive sensors for monitoring core temperature. J Clin Monit Comput. Published 22 March 2025. doi:10.1007/s10877-025-01289-9

Composed by AI from the published abstract and checked against it. How Briefs are made →