Feasibility analysis of a novel non-invasive ultrasonographic method for the measurement of intra-abdominal pressure in the intensive care unit
Authors' conclusion
Our novel ultrasound-based IAP method displayed good correlation and agreement between IAP and IBP at levels up to 15 mmHg and is an excellent solution for quick decision-making in critically ill patients.
Study context
- Population
- Adult medical ICU patients, university hospital74 intra-abdominal pressure (IAP) readings
- Participants
- 21 patients
- Exposure
- Ultrasonographic IAP measurement vs intra-bladder pressure
- Outcome
- Agreement of ultrasonographic IAP with IBP (Bland and Altman's analysis)
Methodological detailssingle-centre, blinded reference operator
- Single-centre
- Blinded reference operator
Main finding
As reported in the abstract
Agreement with IBP (Bland and Altman)
| Group | Bias | Precision |
|---|---|---|
| IAPUS1 | 0.39 | 1.38 |
| IAPUS2 | 0.61 | 1.51 |
Bias
IAPUS10.39
IAPUS20.61
Precision
IAPUS11.38
IAPUS21.51
Observational data: associations, not effects.
Further findings
- Readings per patient
- 3.5 ± 2.5
- Abdominal wall thickness (mm)
- 24.6 ± 13.1
Original publication
Feasibility analysis of a novel non-invasive ultrasonographic method for the measurement of intra-abdominal pressure in the intensive care unit. J Clin Monit Comput. Published 3 May 2023. doi:10. 1007/ s10877-023-01024-2