Feasibility analysis of a novel non-invasive ultrasonographic method for the measurement of intra-abdominal pressure in the intensive care unit

J Clin Monit ComputPublished Prospective observational study

Authors' conclusion

Our novel ultrasound-based IAP method displayed good correlation and agreement between IAP and IBP at levels up to 15 mmHg and is an excellent solution for quick decision-making in critically ill patients.

Study context

Population
Adult medical ICU patients, university hospital74 intra-abdominal pressure (IAP) readings
Participants
21 patients
Exposure
Ultrasonographic IAP measurement vs intra-bladder pressure
Outcome
Agreement of ultrasonographic IAP with IBP (Bland and Altman's analysis)
Methodological detailssingle-centre, blinded reference operator
  • Single-centre
  • Blinded reference operator

Main finding

As reported in the abstract

Agreement with IBP (Bland and Altman)
GroupBiasPrecision
IAPUS10.391.38
IAPUS20.611.51

Bias

IAPUS10.39
IAPUS20.61

Precision

IAPUS11.38
IAPUS21.51

Observational data: associations, not effects.

Further findings

Readings per patient
3.5 ± 2.5
Abdominal wall thickness (mm)
24.6 ± 13.1

Original publication

Feasibility analysis of a novel non-invasive ultrasonographic method for the measurement of intra-abdominal pressure in the intensive care unit. J Clin Monit Comput. Published 3 May 2023. doi:10.1007/s10877-023-01024-2

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.