Safety, quality & education

Patient characteristics and monitor alarms in a cardiac ICU: A retrospective cross-sectional study

J Clin Monit ComputPublished Retrospective cross-sectional study

Authors' conclusion

Routinely available patient characteristics were associated with selected alarm patterns, but the findings are hypothesis-generating.

Study context

Population
Patients in a cardiac surgical ICU, Helsinki University Hospital, over 4 consecutive weeks
Participants
81 patients
Outcome
Patient monitor alarm burden and association of routinely available patient characteristics with alarm frequency, type and duration
Methodological detailssingle-centre, negative-binomial regression
  • Single-centre
  • Negative-binomial regression

Main finding

As reported in the abstract

Alarm burden and associated patient characteristics
Total alarms
45,044
Median alarm rate, alarms per patient per hour [range as printed]
6.8 [4.8 to 10.8]
Older age: fewer alarms
IRR 0.90 (95% CI 0.83 to 0.98, P < 0.05)
ASA 5: more low-ETCO2 alarms
IRR 9.05 (95% CI 3.37 to 27.88, P < 0.05)
Smoking: more low invasive BP alarms
IRR 1.94 (95% CI 1.05 to 3.90, P < 0.05)

Observational data: associations, not effects.

Secondary outcomes

Alarm rates vs two previous studies with identical monitors in other units

Alarm rates similar across 3 clinically different units

Original publication

Patient characteristics and monitor alarms in a cardiac ICU: A retrospective cross-sectional study. J Clin Monit Comput. Published 29 August 2026. doi:10.1007/s10877-026-01486-0

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.