Perioperative medicine & outcomes

Performance of pulse oximeters as a function of race compared to skin pigmentation: a single center retrospective study

J Clin Monit ComputPublished Retrospective cohort study

Authors' conclusion

There was no clinically significant association between error rates and Massey-Martin scale grades and no clinically significant difference in accuracy observed between self-reported Black and White patients.

Study context

Population
Peri-operative patients enrolled in unrelated clinical trials with arterial blood gases (PaO2 ≤ 125 mmHg) and time-matched SpO24,030 paired SaO2-SpO2 values
Participants
579 patients
Exposure
SpO2 accuracy (SaO2-SpO2) by Massey-Martin skin pigmentation score vs self-reported race
Outcome
Incidence of occult hypoxemia and accuracy of pulse oximetry (SaO2-SpO2 error) by Massey-Martin scale
Methodological detailssingle-centre
  • Single-centre

Main finding

As reported in the abstract

Average error (SaO2-SpO2)

Overall average error (SaO2-SpO2) ± SD, %

0.23 ± 2.6%

Range of average errors across Massey scores (statistically significant differences)

- 0.39 ± 2.3 to 0.53 ± 2.5

Range of average errors across races (statistically significant differences)

- 0.55 ± 2.1 to 0.37 ± 2.7

Observational data: associations, not effects.

Further findings

Skin color varied widely within each self-identified race category

Original publication

Performance of pulse oximeters as a function of race compared to skin pigmentation: a single center retrospective study. J Clin Monit Comput. Published 28 August 2024. doi:10.1007/s10877-024-01211-9