Safety, quality & education

Portable handheld ultrasound for venous excess ultrasound (VExUS) assessment in critical care: Reliability and time efficiency in resident-led examinations

J Crit CarePublished Diagnostic agreement and reliability study

Authors' conclusion

Handheld ultrasound provides reliable VExUS assessment comparable to conventional systems (VExUS grade κ > 0.85), with 40 % faster procedural time (5 vs 8 min, p < 0.0001).

Study context

Population
Critically ill ICU patients, VExUS scans by 2 resident trainees after a short VExUS course
Participants
80 patients
Comparison
Handheld vs conventional ultrasound for VExUS assessment
Outcome
Inter-rater reliability and inter-user reproducibility of the VExUS score
Methodological detailsblinded, paired comparison
  • Blinded
  • Paired comparison

Performance

As reported in the abstract

Inter-rater VExUS grading reliability
Test or modelκIntraclass correlation coefficient (ICC)
Handheld0.8790.912
Conventional0.8510.865

κ

Handheld0.879
Conventional0.851

Intraclass correlation coefficient (ICC)

Handheld0.912
Conventional0.865

Further findings

Procedural time

P < 0.0001

OutcomeHandheldConventionalBias (cm)P value
Agreement handheld vs conventional, IVC diameterRho = 0.92, ICC = 0.980.037
Renal vein concordance
κ0.550.653
Grading accuracyComparableP = 0.747

Agreement handheld vs conventional, hepatic velocities

Negligible differences (ICC)
> 0.96

Image acquisition success

Overall
>91 %
Hepatic/IVC
100 %

Original publication