Airway management

Single-use versus reusable metallic laryngoscopes for non-emergent intubation: A retrospective review of 72,672 intubations

J Clin AnesthPublished Retrospective cohort study

Main findingResult — go to the resultAdjusted odds ratio 0.81 95% CI 0.66–0.99

Authors' conclusion

Metallic single-use laryngoscopes were associated with less need for rescue intubation with alternate devices and lower incidence of poor laryngeal view compared to reusable metallic laryngoscopes.

Study context

Population
Adults undergoing non-emergent procedures under GA requiring tracheal intubation, academic medical center
Participants
72,672 adults
Exposure
Metallic single-use vs metallic reusable laryngoscopes
  • Single-usen = 37,123
  • Reusablen = 35,549reference
Outcome
Need for intubation rescue with an alternate device
Methodological detailssingle-site, before-after transition
  • Single-site
  • Before-after transition

Main finding

As reported in the abstract

Adjusted odds ratio

0.81

95% CI 0.66–0.99

Single-use vs Reusable
aOR 0.81 (95% CI 0.66–0.99)

Observational data: associations, not effects.

Secondary outcomes

OutcomeOR (95% CI)
Difficult laryngeal view (modified Cormack-Lehane grade ≥ 2b)0.86 (0.80–0.93)
Hypoxemia (SpO2 < 90% for >30 s) during intubation attempt1.03 (0.88–1.20)

Subgroups: RSI, Macintosh blades, Miller blades, difficult airway risk factors (OSA, Mallampati ≥3, BMI >30 kg/m²) (subgroup)

Similar results observed in subgroup analyses

Original publication

Single-use versus reusable metallic laryngoscopes for non-emergent intubation: A retrospective review of 72,672 intubations. J Clin Anesth. Published 18 June 2023. doi:10.1016/j.jclinane.2023.111187

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.