Transcutaneous carbon dioxide monitoring during prolonged apnoea with high‐flow nasal oxygen

Acta Anaesthesiol ScandPublished Method comparison study

Authors' conclusion

Transcutaneous carbon dioxide monitoring provided a clinically acceptable substitute for arterial blood gases but as hypercapnia developed to considerable levels, we observed overestimation at high carbon dioxide tensions in patients undergoing apnoeic oxygenation with high-flow nasal oxygen.

Study context

Population
Anaesthetised and paralysed patients undergoing prolonged apnoeic oxygenation with HFNO until pH 7.15 or PCO2 12 kPa
Participants
35 participants
Comparison
Transcutaneous CO2 monitoring vs arterial CO2 tension
Primary endpoint
Agreement of transcutaneous CO2 with arterial CO2 tension (Bland-Altman)
Methodological detailssecondary analysis of a study of prolonged apnoeic oxygenation, Bland-Altman analysis
  • Secondary analysis of a study of prolonged apnoeic oxygenation
  • Bland-Altman analysis

Primary result

As reported in the abstract

Agreement of transcutaneous vs arterial CO2

Bias range

−0.55 to 0.81 kPa

Limits of agreement

between −1.25 and 2.11 kPa

Further findings

Apnoea duration and blood gases at termination of apnoeic oxygenation

Median apnoea duration
25 min
Mean pH
7.14
Mean arterial CO2 tension
11.2 kPa

Original publication

Transcutaneous carbon dioxide monitoring during prolonged apnoea with high‐flow nasal oxygen. Acta Anaesthesiol Scand. Published 19 February 2023. doi:10.1111/aas.14216