A Wireless Photoplethysmography Chest Patch for Continuous Vital Sign Monitoring: A Clinical Validation Study in Intensive Care Patients

Acta Anaesthesiol ScandPublished Clinical validation study

Authors' conclusion

Overall agreement was variable, with wide limits of agreement and a substantial proportion of measurements outside predefined thresholds.

Study context

Population
ICU patients1931.8 monitoring hours
Participants
32 patients
Comparison
Chest patch vs Reference monitor
Outcome
Agreement (bias and limits of agreement) between biosensor and reference monitor for five vital signs
Methodological detailsBland-Altman analysis for repeated measurements, Deming regression, Clarke Error Grid analysis
  • Bland-Altman analysis for repeated measurements
  • Deming regression
  • Clarke Error Grid analysis

Performance

As reported in the abstract

Agreement of wearable biosensor with reference bedside monitor (Bland-Altman bias, LoA)
Heart rate bias, beats per minute (LoA −17.06 to 17.38)
0.16
Heart rate, measurements exceeding 10% threshold
18.8%
Heart rate, Clarke Error Grid zones D-E
0.42%
Respiratory rate bias, breaths per minute (LoA −6.78 to 14.31)
3.77
Respiratory rate, measurements exceeding 10% threshold
61.4%
Respiratory rate, Clarke Error Grid zones D-E
18.6%
SBP bias, mmHg
−4.69
Oxygen saturation bias
−1.34%
Temperature bias
−0.29°C

Original publication

A Wireless Photoplethysmography Chest Patch for Continuous Vital Sign Monitoring: A Clinical Validation Study in Intensive Care Patients. Acta Anaesthesiol Scand. Published 26 April 2026. doi:10.1111/aas.70242

Composed by AI from the published abstract and checked against it. How Briefs are made →