Airway management

Assessing a novel second generation laryngeal mask airway using the ‘ADEPT’ approach: results from the LMA® Protector™ observational study

J Clin Monit ComputPublished Prospective observational study

Authors' conclusion

The LMA Protector appears suitable for clinical use

Study context

Population
Adult patients undergoing GA
Participants
111 sequential adults
Outcome
Time to insertion of the device and ability to ventilate after insertion
Methodological detailsADEPT device evaluation framework, sequential patients
  • ADEPT device evaluation framework
  • Sequential patients

Main finding

As reported in the abstract

Insertion time and ability to ventilate

Ability to ventilate after insertion, %

100

95% CI 96.7–100

Observational data: associations, not effects.

Secondary outcomes

≥1 manoeuvre required to assist insertion

Cases requiring ≥1 manoeuvre, %
60.3 (95% CI 50.6-69.5)

Ease of insertion score (5-point scale)

Median ease of insertion score (2-5[3-5])
4

Adequacy of ventilation score (5-point scale)

Median adequacy of ventilation score (5-5[4-5])
5

First-time insertion failure

First-time insertion failure rate
9.9% (95% CI 5.1-17.0%)

Safety

Patient harm, particularly desaturation

No episodes of patient harm recorded

Original publication

Assessing a novel second generation laryngeal mask airway using the ‘ADEPT’ approach: results from the LMA® Protector™ observational study. J Clin Monit Comput. Published 5 September 2022. doi:10.1007/s10877-022-00910-5

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.