Assessing a novel second generation laryngeal mask airway using the ‘ADEPT’ approach: results from the LMA® Protector™ observational study
Authors' conclusion
The LMA Protector appears suitable for clinical use
Study context
- Population
- Adult patients undergoing GA
- Participants
- 111 sequential adults
- Outcome
- Time to insertion of the device and ability to ventilate after insertion
Methodological detailsADEPT device evaluation framework, sequential patients
- ADEPT device evaluation framework
- Sequential patients
Main finding
As reported in the abstract
Insertion time and ability to ventilate
Ability to ventilate after insertion, %
100
95% CI 96.7–100
Observational data: associations, not effects.
Secondary outcomes
≥1 manoeuvre required to assist insertion
- Cases requiring ≥1 manoeuvre, %
- 60.3 (95% CI 50.6-69.5)
Ease of insertion score (5-point scale)
- Median ease of insertion score (2-5[3-5])
- 4
Adequacy of ventilation score (5-point scale)
- Median adequacy of ventilation score (5-5[4-5])
- 5
First-time insertion failure
- First-time insertion failure rate
- 9.9% (95% CI 5.1-17.0%)
Safety
Patient harm, particularly desaturation
- No episodes of patient harm recorded
Original publication
Assessing a novel second generation laryngeal mask airway using the ‘ADEPT’ approach: results from the LMA® Protector™ observational study. J Clin Monit Comput. Published 5 September 2022. doi:10. 1007/ s10877-022-00910-5