Pediatric anesthesia & critical care
Assessment of accuracy of two pulse oximeters in infants with cyanotic and acyanotic congenital heart diseases
Authors' conclusion
In lower range saturations between 70% and 85% the accuracy of both pulse oximeters exceeded the threshold of ≤3% recommended by the Food and Drug Administration (FDA).
Study context
- Population
- Infants with congenital heart diseases, arterial oxygen saturation 70–100%153 measurements
- Participants
- 44 infants
- Comparison
- Philips IntelliVue X3 vs Nellcor OxiMax pulse oximeters, each vs arterial saturation (SaO2) by co-oximetry
- Outcome
- Accuracy of SpO2 vs SaO2 (arterial blood gas analysis): median bias and accuracy root mean squared index (Arms), by SaO2 subgroup
Performance
As reported in the abstract
Median bias (SpO2 − SaO2) and Arms by SaO2 subgroup (group 1 70.0%-85.0%, group 2 85.1%-94.0%, group 3 >94.1%)
Arms group 1
Arms group 2
Arms group 3
| Outcome | Philipps | OxiMax™ |
|---|---|---|
| Median bias group 1 (IQR 2.6-8.7) | 5.3% | — not stated in the abstract |
| Median bias group 2 (IQR 0.9-6.0) | 2.3% | — not stated in the abstract |
| Median bias group 3 (IQR −0.8-2.4) | 1.1% | — not stated in the abstract |
| Median bias group 1 (IQR 0.5-5.1) | — not stated in the abstract | 2.7% |
| Median bias group 2 (IQR −0.9-2.6) | — not stated in the abstract | 0.2% |
| Median bias group 3 (IQR −1.3-0.6) | — not stated in the abstract | −0.5% |
— not stated in the abstract
Original publication
Assessment of accuracy of two pulse oximeters in infants with cyanotic and acyanotic congenital heart diseases. Paediatr Anaesth. Published 4 October 2024. doi:10. 1111/ pan. 15017
Composed by AI from the published abstract and checked against it. How Briefs are made →