Assessment of accuracy of two pulse oximeters in infants with cyanotic and acyanotic congenital heart diseases

Paediatr AnaesthPublished Diagnostic accuracy study

Authors' conclusion

In lower range saturations between 70% and 85% the accuracy of both pulse oximeters exceeded the threshold of ≤3% recommended by the Food and Drug Administration (FDA).

Study context

Population
Infants with congenital heart diseases, arterial oxygen saturation 70–100%153 measurements
Participants
44 infants
Comparison
Philips IntelliVue X3 vs Nellcor OxiMax pulse oximeters, each vs arterial saturation (SaO2) by co-oximetry
Outcome
Accuracy of SpO2 vs SaO2 (arterial blood gas analysis): median bias and accuracy root mean squared index (Arms), by SaO2 subgroup

Performance

As reported in the abstract

Median bias (SpO2 − SaO2) and Arms by SaO2 subgroup (group 1 70.0%-85.0%, group 2 85.1%-94.0%, group 3 >94.1%)

Arms group 1

Arms group 2

Arms group 3

OutcomePhilippsOxiMax™
Median bias group 1 (IQR 2.6-8.7)5.3%— not stated in the abstract
Median bias group 2 (IQR 0.9-6.0)2.3%— not stated in the abstract
Median bias group 3 (IQR −0.8-2.4)1.1%— not stated in the abstract
Median bias group 1 (IQR 0.5-5.1)— not stated in the abstract2.7%
Median bias group 2 (IQR −0.9-2.6)— not stated in the abstract0.2%
Median bias group 3 (IQR −1.3-0.6)— not stated in the abstract−0.5%

— not stated in the abstract

Original publication

Assessment of accuracy of two pulse oximeters in infants with cyanotic and acyanotic congenital heart diseases. Paediatr Anaesth. Published 4 October 2024. doi:10.1111/pan.15017