Cardiothoracic & vascular anesthesia

Cardiac output monitoring using an automatic arm cuff device in surgical patients

Br J AnaesthPublished Method comparison study

Authors' conclusion

These initial findings suggest potential for convenient CO monitoring using an automatic arm cuff device in surgical patients.

Study context

Population
Liver transplant and cardiac surgery patientsfrom 24 liver transplant and 10 cardiac surgery patients · liver transplant 24 · cardiac surgery 10
Participants
191 measurement sets participants
Comparison
Arm cuff-based cardiac output (CO) and pulse contour analysis against reference CO from pulmonary artery catheter bolus thermodilution
Outcome
Agreement with reference CO

Main finding

As reported in the abstract

GroupCorrelationConcordance rate
Arm cuff-based method0.61 (95% confidence interval 0.37-0.79)83% (75%-91%)
Classic pulse contour analysis0.59 (0.37-0.75)81% (72%-90%)

Correlation

Arm cuff-based method0.61 (95% confidence interval 0.37-0.79)
Classic pulse contour analysis0.59 (0.37-0.75)

Concordance rate

Arm cuff-based method83% (75%-91%)
Classic pulse contour analysis81% (72%-90%)

Original publication

Cardiac output monitoring using an automatic arm cuff device in surgical patients. Br J Anaesth. Published 5 November 2026. doi:10.1016/j.bja.2026.09.020

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.