Reliability of Bioreactance and Pulse-Power Analysis in Measuring Cardiac Index During Open Abdominal Aortic Surgery

J Cardiothorac Vasc AnesthPublished Method-comparison study

Authors' conclusion

The CI measurements achieved with Starling SV and LiDCOrapid were not interchangeable with TDCO, and the ability to track changes in CI was poor.

Study context

Population
Patients undergoing elective open abdominal aortic surgery, Oulu University Hospital, Finland627 and 497 measurement pairs
Participants
40 patients
Exposure
Bioreactance (Starling SV) and pulse-power analysis (LiDCOrapid) vs pulmonary artery catheter thermodilution (thermodilution cardiac output (TDCO))
Outcome
Accuracy, precision and trending ability of cardiac index vs TDCO (Bland-Altman agreement; four-quadrant plots with error grids)
Follow-up
Intraoperative
Methodological detailsprospective, single-centre
  • Prospective
  • Single-centre

Main finding

As reported in the abstract

Agreement with TDCO (Bland-Altman)

Bias

0.18 L/min/m²

95% CI 0.13–0.23

Starling SV · wide LOA −1.12 to 1.47 L/min/m²

Bias

−0.15 L/min/m²

95% CI −0.21 to −0.09

LiDCOrapid · wide LOA −1.56 to 1.37

Starling SV

63.7

95% CI = 52.4–71.0

Percentage error (PE)

LiDCOrapid

68.7

95% CI = 54.9–79.6

PE

Observational data: associations, not effects.

Further findings

Trending ability (four-quadrant plots with error grids)

Trending ability of neither device was sufficient

Original publication

Reliability of Bioreactance and Pulse-Power Analysis in Measuring Cardiac Index During Open Abdominal Aortic Surgery. J Cardiothorac Vasc Anesth. Published 12 February 2024. doi:10.1053/j.jvca.2024.02.005

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.