Accuracy of Cardiac Output Measured by Fourth-Generation FloTrac and LiDCOrapid, and Their Characteristics Regarding Systemic Vascular Resistance in Patients Undergoing Cardiac Surgery

J Cardiothorac Vasc AnesthPublished Prospective observational study

Authors' conclusion

The accuracy of CIFT or CILR was not clinically acceptable for cardiac surgery. Fourth-generation FloTrac was unreliable in high SVRI. LiDCOrapid was inaccurate across a broad range of SVRI, and minimally affected by SVRI.

Study context

Population
Adults undergoing elective cardiac surgery, single university hospital135 measurements
Participants
29 adults
Exposure
Fourth-generation FloTrac (CIFT) and LiDCOrapid (CILR) vs intermittent thermodilution via pulmonary artery catheter (CITD)
Outcome
Agreement with CITD
Methodological detailssingle-centre
  • Single-centre

Main finding

As reported in the abstract

OutcomeCIFTCILR
R0.620.58
Bias−0.73 L/min/m²−0.61 L/min/m²
Limit of agreement−2.14-to-0.68 L/min/m²−2.42-to-1.20 L/min/m²
Percentage error39.9%51.2%

Observational data: associations, not effects.

Further findings

OutcomeCIFTCILR
Percentage error by SVRI (subgroup)
Low SVRI (<1,200 dyne×s/cm5/m)33.9%54.5%
Moderate SVRI (1,200-1,800 dyne×s/cm5/m)37.6%47.9%
High SVRI (>1,800 dyne·s/cm5/m²)49.3%50.6%

Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.

Original publication

Accuracy of Cardiac Output Measured by Fourth-Generation FloTrac and LiDCOrapid, and Their Characteristics Regarding Systemic Vascular Resistance in Patients Undergoing Cardiac Surgery. J Cardiothorac Vasc Anesth. Published 17 March 2023. doi:10.1053/j.jvca.2023.03.019

Composed by AI from the published abstract and checked against it. How Briefs are made →