Cardiothoracic & vascular anesthesia

Electrical cardiometry for non-invasive cardiac output monitoring: a method comparison study in patients after coronary artery bypass graft surgery

J Clin Monit ComputPublished Method comparison study

Main finding · CO agreementResult — go to the resultMean of the differences −0.2 L/min 95% CI −0.5 to 0.2

Authors' conclusion

In this study, the agreement between EC-CO and PATD-CO was not clinically acceptable in patients after CABG surgery. The trending ability of EC-CO was poor.

Study context

Population
Patients after coronary artery bypass graft (CABG) surgery157 paired measurements
Participants
41 patients
Exposure
PATD-CO vs electrical cardiometry (EC-CO)
Outcome
Agreement of CO measurements (mean of differences with 95%-limits of agreement, Bland-Altman; percentage error)
Methodological detailsprospective, Bland-Altman analysis, four-quadrant plot analysis
  • Prospective
  • Bland-Altman analysis
  • Four-quadrant plot analysis

Main finding

CO agreement

Mean of the differences

−0.2 L/min

95% CI −0.5 to 0.2

By group

PATD-CO vs EC-CO
mean of the differences −0.2 L/min (95% CI −0.5 to 0.2)

As reported in the abstract

Percentage error
47% (95%-CI 37 to 56%)

Observational data: associations, not effects.

Further findings

Trending ability (four-quadrant plot)

Concordance rate for CO changes
48%

Original publication

Electrical cardiometry for non-invasive cardiac output monitoring: a method comparison study in patients after coronary artery bypass graft surgery. J Clin Monit Comput. Published 11 December 2024. doi:10.1007/s10877-024-01246-y

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.