The Validation of Cardiac Index and Stroke-Volume Variation Measured by the Pulse-Wave Transit Time-Analysis Versus Conventional Pulse-Contour Analysis After Off-Pump Coronary Artery Bypass Grafting: Observational Study
Authors' conclusion
The overall performance of CIesCCO and esSVV versus CIPCA and SVVPCA is not clinically acceptable.
Study context
- Population
- Patients after elective off-pump CABG (OPCAB), 1,000-bed university hospital
- Participants
- 21 patients · 178 CI pairs · 174 stroke-volume variation (SVV) pairs
- Exposure
- CI and SVV by pulse-wave transit-time (PWTT) vs conventional pulse-contour analysis
- Outcome
- Agreement esCCO vs PCA
- Follow-up
- 10 study stages
Methodological detailssingle-center, prospective, observational
- Single-center
- Prospective
- Observational
Main finding
As reported in the abstract
- CI mean bias (CIesCCO vs CIPCA)
- 0.06 L min/m²
- CI limits of agreement
- ± 0.92 L min/m²
- CI percentage error
- 35.3%
- SVV mean bias (esSVV vs SVVPCA)
- −6.1%
- SVV limits of agreement
- ± 15.5%
- SVV percentage error
- 137%
Observational data: associations, not effects.
Further findings
Trending ability of CIesCCO vs CIPCA
- Concordance rate
- 70%
Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.
Original publication
The Validation of Cardiac Index and Stroke-Volume Variation Measured by the Pulse-Wave Transit Time-Analysis Versus Conventional Pulse-Contour Analysis After Off-Pump Coronary Artery Bypass Grafting: Observational Study. J Cardiothorac Vasc Anesth. Published 11 February 2023. doi:10. 1053/ j. jvca. 2023. 02. 010