Respiratory failure & ventilation

Central Venous Pressure Referencing in the Lateral Position: Comparison With Direct Right Atrial Pressure in Mechanically Ventilated ICU Patients

Acta Anaesthesiol ScandPublished Prospective observational study

Authors' conclusion

In mechanically ventilated ICU patients, central venous pressure accuracy was position-dependent, and optimal external reference levels differed between lateral positions. Right atrial pressure increased in the right lateral position, suggesting that lateral positioning influences cardiopulmonary physiology beyond hydrostatic measurement effects alone.

Study context

Population
Sedated, mechanically ventilated ICU patients with central venous cathetersevery patient in every position
Participants
16 participants
Exposure
Central venous pressure (CVP) (transducer 5 cm below mid-sternum) vs direct right atrial pressure (RAP) (solid-state catheter, mid-right atrium) in supine, 45° left lateral and 45° right lateral positions
Outcome
Agreement between RAP and CVP (Bland-Altman bias and limits of agreement, ICC, predefined clinical thresholds)
Follow-up
End-expiration
Methodological detailsmethod-comparison, Bland-Altman analysis
  • Method-comparison
  • Bland-Altman analysis

Main finding

As reported in the abstract

Agreement of CVP (5 cm below sternum) with RAP
GroupValueICC
Right lateralbias −0.4 (limits of agreement −3.2 to +2.4) mmHg0.85
Left lateralbias −2.2 (limits of agreement −5.0 to +0.7) mmHg0.58

Value

Right lateralbias −0.4 (limits of agreement −3.2 to +2.4) mmHg
Left lateralbias −2.2 (limits of agreement −5.0 to +0.7) mmHg

ICC

Right lateral0.85
Left lateral0.58

Observational data: associations, not effects.

Further findings

OutcomeΔ (mmHg)P value
RAP change with positionRight lateral vs Supine · RAP increased in right lateral+2.9P < 0.001
RAP change with positionRight lateral vs Left lateral · RAP increased in right lateral+2.8P < 0.001
RAP change with positionLeft lateral vs Supine · no differenceP = 0.78

Optimal CVP reference level, left lateral position (hydrostatic adjustment)

Bias with adjustment to ≈2 cm below the sternum (minimised)
−0.1 mmHg
ICC with adjustment to ≈2 cm below the sternum (improved agreement)
0.82

Original publication

Central Venous Pressure Referencing in the Lateral Position: Comparison With Direct Right Atrial Pressure in Mechanically Ventilated ICU Patients. Acta Anaesthesiol Scand. Published 12 June 2026. doi:10.1111/aas.70280

Composed by AI from the published abstract and checked against it. How Briefs are made →