Obstetric anesthesia

Comparison between the Rotational Thromboelastometry (ROTEM) Delta device against the Cartridge-based Thromboelastography 6s and Quantra in a healthy third trimester pregnant cohort

J Clin Monit ComputPublished Method comparison study

Authors' conclusion

All the POCVT devices demonstrated a high correlation to laboratory Clauss fibrinogen, making each beneficial for the early recognition and management of hypofibrinogenemia.

Study context

Population
Healthy third trimester pregnant women
Comparison
ROTEM Delta vs TEG 6s and Quantra
Outcome
Correlation of 3 POCVT devices with laboratory Clauss fibrinogen and platelets

Main finding

As reported in the abstract

Correlation with Clauss fibrinogen
GroupValueP
ROTEMr 0.76-0.84P < 0.0001
TEG6sr 0.71P < 0.0001
Quantrar 0.72P = 0.0001

Value

ROTEMr 0.76-0.84
TEG6sr 0.71
Quantrar 0.72

P

ROTEMP < 0.0001
TEG6sP < 0.0001
QuantraP = 0.0001

Further findings

OutcomeROTEMQuantraEstimateP value
Correlation with laboratory plateletsModerate correlationr 0.54;0.45r 0.66
PP < 0.0001; P = 0.0013P = 0.0001
Inter-device correlation, fibrinogen parametersTEG6s vs ROTEM · highr 0.88P < 0.0001
Inter-device correlation, fibrinogen parametersQuantra vs ROTEM · highr 0.74P < 0.0001
Inter-device correlation, platelet parametersQuantra vs ROTEM · moderater 0.51P = 0.0036
Total running timeQuantra vs ROTEM · if warming and pipetting of ROTEM reagents considereddifference (faster) 20.9 min (95% CI −0.2 to 4.7)P = 0.07

Original publication

Comparison between the Rotational Thromboelastometry (ROTEM) Delta device against the Cartridge-based Thromboelastography 6s and Quantra in a healthy third trimester pregnant cohort. J Clin Monit Comput. Published 29 July 2022. doi:10.1007/s10877-022-00888-0