Comparison of a Modern Digital Mechanomyograph to a Mechanomyograph Utilizing an Archival Grass Force Transducer

AnesthesiologyPublished Method comparison study

Authors' conclusion

The new mechanomyograph resulted in similar train-of-four ratio measurements compared to an archival mechanomyography system utilizing an FT-10 Grass force transducer.

Study context

Population
Patients undergoing surgery with rocuronium NMB767 paired TOF measurements
Participants
eight patients
Comparison
New mechanomyograph vs Archival system
Outcome
Equivalence of TOF ratios recorded on each device
Follow-up
Onset and recovery from rocuronium NMB
Methodological detailspaired measurement on opposite arms
  • Paired measurement on opposite arms

Main finding

As reported in the abstract

TOF ratio agreement

Linear association of paired TOF ratios (R 2)

0.94

P < 0.001

Bland-Altman bias

3.8%

Bland-Altman limits of agreement

−13% and 21%

Further findings

OutcomeNew mechanomyographArchival systemP value
Precision and measurement sensitivityFor both precision and sensitivityP < 0.001
Precision5.3 mV15.5 mV
Measurement sensitivity1.6 mV5.7 mV

Calibration linearity (correlation coefficient)

Linear correlation coefficient for both devices
1.00

Original publication

Comparison of a Modern Digital Mechanomyograph to a Mechanomyograph Utilizing an Archival Grass Force Transducer. Anesthesiology. Published 17 April 2023. doi:10.1097/aln.0000000000004590

Composed by AI from the published abstract and checked against it. How Briefs are made →