Perioperative medicine & outcomes

Continuous vital sign monitoring with Biobeat® wearable devices for post-ambulatory surgery patients: a pilot feasibility study

J Clin Monit ComputPublished Prospective observational feasibility study

Authors' conclusion

Continuous vital signs monitoring is feasible for high-risk outpatient surgery patients; however, significant improvements are required in device reliability and data accessibility.

Study context

Population
Post-ambulatory (outpatient) surgery patients, CHUM, Montréal, February–August 202349 completed
Participants
50 participants
Outcome
Feasibility: percentage of patients without data loss during consecutive 2-hour intervals
Methodological detailspilot, single-centre
  • Pilot
  • Single-centre

Main finding

As reported in the abstract

Patients without data loss

Patients maintaining connectivity without complete data loss for 6-8-hour intervals

39.6%

Observational data: associations, not effects.

Further findings

Patients with postoperative complications
3
No data available within 24 h prior to the episodes

Safety

Early device removal

Patients removing the device early due to discomfort or technical issues
13
  • Challenges: device discomfort, Bluetooth disconnection, connectivity issues

Original publication

Continuous vital sign monitoring with Biobeat® wearable devices for post-ambulatory surgery patients: a pilot feasibility study. J Clin Monit Comput. Published 26 February 2025. doi:10.1007/s10877-025-01276-0

Composed by AI from the published abstract and checked against it. How Briefs are made →