Pediatric anesthesia & critical care
Feasibility of Obtaining Pulse Oximetry Readings From the Oropharynx
Authors' conclusion
Our preliminary data shows that oximetry values that are clinically in-line with extremity oximetry values can be obtained from the oropharynx. The accuracy appears to be greater with the emitter and sensor facing the tongue, although there was a higher percentage of missed values.
Study context
- Participants
- 50 vs 50
- Comparison
- Cohort 1 vs Cohort 2
- Cohort 1n = 50
- Cohort 2n = 50
- Outcome
- Feasibility of obtaining pulse oximetry measurements from the oropharynx
Methodological detailstwo cohorts
- Two cohorts
Main finding
As reported in the abstract
Absolute difference from extremity oximetry values
Cohort 1
4.2
± 3.4
Cohort 2
2.5
± 3.4
Further findings
Values within 0%-2% of extremity oximetry value
Fallout (oropharyngeal oximeter did not record a value)
Original publication
Feasibility of Obtaining Pulse Oximetry Readings From the Oropharynx. Paediatr Anaesth. Published 31 July 2025. doi:10. 1111/ pan. 70023