Respiratory failure & ventilation

Low Perfusion and Missed Diagnosis of Hypoxemia by Pulse Oximetry in Darkly Pigmented Skin: A Prospective Study

Anesth AnalgPublished Prospective laboratory study

Authors' conclusion

Low peripheral perfusion combined with darker skin pigmentation leads to clinically significant high-reading pulse oximeter errors and missed diagnoses of hypoxemia.

Study context

Population
Healthy subjects during stable hypoxemia (SaO2 68%-100%), 2 pulse oximeters (Nellcor N-595, Masimo Radical 7)9,763 matched readings
Participants
146 subjects
Exposure
Light vs Medium vs Dark
  • Lightn = 25reference
  • Mediumn = 78
  • Darkn = 43
Outcome
Pulse oximeter bias (difference between SaO2 and SpO2) by skin pigment category, in a multivariable mixed-effects model incorporating repeated measures, SaO2 levels and perfusion
Methodological detailslaboratory setting, multivariable mixed-effects model
  • Laboratory setting
  • Multivariable mixed-effects model

Main finding

As reported in the abstract

Pulse oximeter bias
Perfusion index (PI) 1.0% to 1.5%, 0.5% to 1.0% and <0.5%: relationship to mean bias or median absolute bias
P<.00001

Observational data: associations, not effects.

Further findings

Missed diagnosis of hypoxemia (SpO2 92%-96% when SaO2 <88%) in low perfusion (PI < 1%) · combined frequency, both oximeters

Original publication

Low Perfusion and Missed Diagnosis of Hypoxemia by Pulse Oximetry in Darkly Pigmented Skin: A Prospective Study. Anesth Analg. Published 18 December 2023. doi:10.1213/ane.0000000000006755

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.