P0.1 is an Unreliable Measure of Effort in Support Mechanical Ventilation in Comparison With Esophageal-Derived Measures of Effort: A Comparison Study

Crit Care MedPublished Retrospective observational comparison study

Authors' conclusion

The results demonstrated poor correlations between P0.1 and the other effort parameters, whereas strong correlations were observed among ΔPes, WOB, and PTP.

Study context

Population
Mechanically ventilated patients in spontaneous breathing (support) mode, mixed medical-surgical ICU, Leiden University Medical Center117 timeframes from 39 patient cases analysed
Participants
30 participants
Exposure
P0.1 vs esophageal-derived measures of effort
Outcome
Correlation of P0.1 with esophageal-derived effort parameters
Methodological detailssingle-centre, Average values over at least three 5-minute time frames per patient
  • Single-centre
  • Average values over at least three 5-minute time frames per patient

Main finding

As reported in the abstract

P0.1 vs work of breathing (WOB) (poor correlation), R2

0.111

P0.1 vs pressure-time product (PTP) (poor correlation), R2

0.113

P0.1 vs ΔPes (poor correlation), R2

0.034

Observational data: associations, not effects.

Further findings

Correlations among esophageal-derived parameters

PTP vs WOB (high correlation), R2
0.886
ΔPes vs WOB (high correlation), R2
0.848
ΔPes vs PTP (high correlation), R2
0.876

Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.

Original publication

P0.1 is an Unreliable Measure of Effort in Support Mechanical Ventilation in Comparison With Esophageal-Derived Measures of Effort: A Comparison Study. Crit Care Med. Published 11 June 2025. doi:10.1097/ccm.0000000000006745

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.