Respiratory failure & ventilation

Prehospital Normoventilation in Critically Ill Anaesthetised Patients Cannot Be Reliably Monitored by Capnography—A Retrospective Observational Study

Acta Anaesthesiol ScandPublished Retrospective observational study

Authors' conclusion

Our results show that for critically ill patients intubated in the prehospital setting, capnography is not accurate in guiding ventilation.

Study context

Population
Anaesthetised and intubated prehospital patients with measured PaCO2, single helicopter EMS unit, 2014–2021
Participants
670 participants
Exposure
Agreement between EtCO2 and PaCO2 across diagnostic groups
  • OHCAn = 221
  • Trauman = 104
  • Neurologicaln = 197
  • Intoxicationn = 99
  • Othern = 49
Outcome
Agreement between EtCO2 and PaCO2 (Bland-Altman); proportion with PaCO2-EtCO2 gradient >0.67 kPa
Methodological detailssingle helicopter EMS unit
  • Single helicopter EMS unit

Main finding

As reported in the abstract

PaCO2-EtCO2 gradient >0.67 kPa
Overall gradient >0.67 kPa: 530/670
79.1% (95% CI 75.9-82.0)
Overall Bland-Altman bias (SD 1.6; limits of agreement −1.4 to 4.8 kPa)
1.7 kPa

Observational data: associations, not effects.

Original publication

Prehospital Normoventilation in Critically Ill Anaesthetised Patients Cannot Be Reliably Monitored by Capnography—A Retrospective Observational Study. Acta Anaesthesiol Scand. Published 29 July 2026. doi:10.1111/aas.70316

Composed by AI from the published abstract and checked against it. How Briefs are made →