Anesthetic drugs & techniques

The impact of electrosurgical devices on electromyography-based neuromuscular monitoring during abdominal laparotomy: a case series

J Clin Monit ComputPublished Prospective observational study

Authors' conclusion

This prospective observational study demonstrated that the use of EMG-based neuromuscular monitoring does not seem to be heavily affected by electrical interference during lower abdominal laparotomic surgery.

Study context

Population
Women aged 32–64 years undergoing gynecological laparotomy under TIVA
Participants
17 women
Outcome
Ratio of measurement failure
Methodological detailscase series
  • Case series

Main finding

As reported in the abstract

Measurement failure ratio, median (range)

3.5%

1.4–6.5%

Observational data: associations, not effects.

Secondary outcomes

Total number of measurements

Total measurements, median (range)
3091 (1480-8134)

Number of measurement failures

Measurement failures, median (range)
94 (60-200)

Most extended consecutive number of measurement failures

Longest run of consecutive failures, median (range)
8 (4-13)

Maintenance and reversal of NMB under electromyography (EMG) guidance

All attending anesthesiologists able to maintain and reverse NMB under EMG guidance

Original publication