Validation of a Novel Method for Noninvasive Mixed Venous Oxygen Saturation Monitoring in Anesthetized Children

Anesth AnalgPublished Method comparison study

Authors' conclusion

In conclusion, this first clinical application of a novel concept for noninvasive SvO 2 monitoring without the need for a pulmonary artery catheter indicates that Capno-SvO 2 generates absolute values and trending capacity in close agreement with the gold standard reference method.

Study context

Population
Children undergoing cardiac catheter interventions, subjected to moderate hemodynamic challenges
Participants
25 procedures
Exposure
Capno-SvO2 vs pulmonary artery blood gas CO-oximetry
Outcome
Agreement of absolute SvO2 values (Bland-Altman) and trending ability (concordance rate)
Methodological detailsprospective, observational
  • Prospective
  • Observational

Main finding

As reported in the abstract

Agreement of Capno-SvO 2 with CO-oximetry (Bland-Altman) and trending ability (concordance rate)

Bias

+3 percentage points

Capno-SvO 2

Upper limit of agreement

+11 percentage points

95% CI 9–14

Capno-SvO 2

Lower limit of agreement

−5 percentage points

95% CI −8 to −3

Capno-SvO 2

Concordance rate (ability to track change)
92% (95% CI, 89-96)

Observational data: associations, not effects.

Original publication

Validation of a Novel Method for Noninvasive Mixed Venous Oxygen Saturation Monitoring in Anesthetized Children. Anesth Analg. Published 2 August 2024. doi:10.1213/ane.0000000000007083

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.