Anesthetic drugs & techniques

A single-centre, randomised comparison of the performance and safety of a novel mechanomyography sensor with electromyography

Br J AnaesthPublished Prospective interventional agreement study

Authors' conclusion

The TOF3D mechanosensor was not less precise than electromyography during baseline.

Study context

Population
Anaesthetised adult patients receiving rocuronium 0.45 mg kg-1 i.v. then sugammadex 2 mg kg-1 i.v.
Participants
33 devices on opposite arms
Exposure
Mechanomyography vs electromyography
Outcome
Precision (repeatability coefficient) of TOF ratio measurement; noninferior if repeatability coefficient did not exceed a margin of 0.01
Follow-up
Baseline and full recovery
Methodological detailssingle-centre, randomised device placement, noninferiority
  • Single-centre
  • Randomised device placement
  • Noninferiority

Main finding

As reported in the abstract

TOF ratio precision
Mechanomyography vs ElectromyographyMedian difference (95% CI)
Baseline; covered the acceptable margin
0.039(0.007–0.042)
After recovery to TOF ratio >0.9; did not cover the margin
0.052(0.047–0.076)

Non-inferiority margin 0.01 as printed, not drawn

Observational data: associations, not effects.

Original publication

A single-centre, randomised comparison of the performance and safety of a novel mechanomyography sensor with electromyography. Br J Anaesth. Published 7 January 2026. doi:10.1016/j.bja.2025.10.067