Safety, quality & education

A smartphone application for semi-automated QT interval analysis based on a snapshot of an electrocardiogram trace displayed on a patient monitor

J Clin Monit ComputPublished Method comparison study

Authors' conclusion

The agreement between QTAPP and QTREF and between QTcAPP and QTcREF was clinically acceptable in adult volunteers and patients without cardiac comorbidities.

Study context

Population
Adult volunteers and patients with sinus rhythm and no acute or chronic cardiac comorbidities
Participants
57 volunteers
Exposure
SMART-QT smartphone app vs manual 12-lead electrocardiogram (ECG) measurement
Outcome
Agreement of QT and QTc intervals (Bland-Altman: mean of differences, SD, 95%-limits of agreement; clinically acceptable if ≤ 20 ± 20 ms)
Methodological detailsprospective, single-centre
  • Prospective
  • Single-centre

Main finding

As reported in the abstract

Agreement (Bland-Altman)
Mean of the differences, ms (95%-LOA CI)
QTAPP vs QTREFSD 20 ms
14(−26 to 54)
QTcAPP vs QTcREFSD 15 ms
13(−16 to 42)

Observational data: associations, not effects.

Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.

Original publication

A smartphone application for semi-automated QT interval analysis based on a snapshot of an electrocardiogram trace displayed on a patient monitor. J Clin Monit Comput. Published 10 April 2025. doi:10.1007/s10877-025-01277-z

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.