Comparison of manual and automated respiratory rate measurements on hospital wards: a prospective observational study

J Clin Monit ComputPublished Prospective observational study

Main finding · Agreement, automated vs time-matched manual respiratory rateResult — go to the resultMean difference 2.5 breaths/minute 95% CI 2.2–2.8

Authors' conclusion

Automated respiratory rate measurements using RespiraSense do not display clinically acceptable agreement with manual measurements in the setting of a respiratory ward.

Study context

Population
Patients on general respiratory wards
Participants
31 patients
Exposure
Automated vs manual respiratory rate measurement
Outcome
Concordance between manual and automated respiratory rate measurements
Methodological detailsmethod comparison, three general respiratory wards
  • Method comparison
  • Three general respiratory wards

Main finding

As reported in the abstract

Agreement, automated vs time-matched manual respiratory rate

Mean difference

2.5 breaths/minute

95% CI 2.2–2.8

Automated vs Manual
mean difference 2.5 breaths/minute (95% CI 2.2–2.8)
Trends in manual and automated respiratory rates concordant
56%

Observational data: associations, not effects.

Further findings

Distribution of readings

Manual readings showed large peaks at 20 and 24 breaths/min; automated readings followed a smooth bell-shaped distribution

Day vs night respiratory rate

Both higher during the day than at night, more marked for automated readings

Original publication

Comparison of manual and automated respiratory rate measurements on hospital wards: a prospective observational study. J Clin Monit Comput. Published 15 November 2025. doi:10.1007/s10877-025-01380-1