Obstetric anesthesia

Agreement between three noninvasive temperature monitoring devices during spinal anaesthesia for caesarean delivery: a prospective observational study

J Clin Monit ComputPublished Prospective observational study

Authors' conclusion

While monitoring of core temperature during obstetric spinal anaesthesia is clinically important, agreement between monitors was below clinically acceptable limits.

Study context

Population
Women undergoing elective and emergency caesarean delivery under spinal anaesthesiaincluded in analysis
Participants
166 women
Exposure
Oral thermometer and infrared thermometer vs heat flux core temperature monitor (Dräger T-core©)
Outcome
Agreement between devices (Bland-Altman LOA; predetermined clinically relevant LOA ± 0.5 °C)
Methodological detailsmethod-comparison (Bland-Altman), error grid analysis
  • Method-comparison (Bland-Altman)
  • Error grid analysis

Main finding

As reported in the abstract

Agreement between devices (Bland-Altman LOA)

Limits of agreement (LOA)

1.8 °C

CI 1.7–2.0

Heat flux vs Oral thermometer · bias 0.5 °C

LOA

2.3 °C

CI 2.1–2.4

Heat flux vs Infrared thermometer · bias 0.4 °C

LOA

2.0 °C

CI 1.9–2.2

Infrared thermometer vs Oral thermometer · bias 0.1 °C

Observational data: associations, not effects.

Further findings

Hypothermia (heat flux temperature < 36 °C)

Incidence of hypothermia
67% (95% CI 49 to 78%)

Error grid analysis

Large amount of clinical disagreement between methods

Original publication

Agreement between three noninvasive temperature monitoring devices during spinal anaesthesia for caesarean delivery: a prospective observational study. J Clin Monit Comput. Published 30 April 2024. doi:10.1007/s10877-024-01154-1

Composed by AI from the published abstract and checked against it. How Briefs are made →