Pediatric anesthesia & critical care

Application and Reliability Evaluation of Sternal Notch Sagittal Ultrasound Imaging for Endotracheal Tube Positioning in Neonates Under General Anesthesia: A Prospective Study Comparing Electronic Bronchoscopy

J Cardiothorac Vasc AnesthPublished Prospective observational study

Authors' conclusion

Sternal notch sagittal ultrasound imaging shows highly consistent measurement results with electronic bronchoscopy for assessing the position of the endotracheal tube in neonates.

Study context

Population
Neonates intubated under GA for elective and urgent surgery in the operating room, Anhui Provincial Children's Hospital, Hefei, China
Exposure
Ultrasound vs Electronic bronchoscopy
Outcome
Agreement between ultrasound-measured ETT tip to right pulmonary artery distance and bronchoscopy-measured ETT tip to carina distance
Methodological detailssingle-center, self-controlled design
  • Single-center
  • Self-controlled design

Main finding

As reported in the abstract

Correlation of US-measured ETT tip–right pulmonary artery distance with bronchoscopy-measured ETT tip–carina distance
High linear correlation between the two methods (r)
0.899
P value for correlation
p < 0.001
Internal consistency of three US measurements (Cronbach alpha)
0.985
Internal consistency of three US measurements (ICC)
0.985

Observational data: associations, not effects.

Further findings

Operation time · significant difference

P < 0.001

Internal consistency of three ultrasound measurements
0.985 · Cronbach alpha · ICC 0.985

Ultrasound precision by ETT size (mean absolute deviation (MAD); coefficient of variation)

2.5-mm ETT: MAD 0.18 mm, CV 5.60%
3.0-mm ETT: MAD 0.14 mm, CV 4.50%
3.5-mm ETT: MAD 0.12 mm, CV 4.20%
  • Improved consistency with larger tube diameters

Safety

OutcomeUltrasoundElectronic bronchoscopyP value
Adverse events / hypoxemia during examinationHypoxemia incidence with bronchoscopy0 events18.4%P < 0.001

Original publication

Application and Reliability Evaluation of Sternal Notch Sagittal Ultrasound Imaging for Endotracheal Tube Positioning in Neonates Under General Anesthesia: A Prospective Study Comparing Electronic Bronchoscopy. J Cardiothorac Vasc Anesth. Published 17 March 2025. doi:10.1053/j.jvca.2025.03.014

Composed by AI from the published abstract and checked against it. How Briefs are made →