Sepsis & infection

Cardiac output monitors in septic shock: do they deliver what matters? A systematic review and meta-analysis

Crit CarePublished Systematic review and meta-analysis of prospective method-comparison studies

Authors' conclusion

Most CO monitors demonstrate poor agreement with reference methods in septic shock.

Study context

Population
Adult ICU patients with septic shock
Participants
1,323 patients · 26 studies
Comparison
Cardiac output (CO) monitoring devices vs reference methods
  • Pulmonary artery catheter (PAC)n = 18 datasetsreference
  • Transpulmonary thermodilution (TPTD)n = 16reference
  • Echocardiographyn = 3reference
Outcome
Agreement with reference methods (pooled bias, LoA, PE; agreement defined as PE < 30%)
Methodological detailsSidik-Jonkman random-effects model, Bland-Altman analysis
  • Sidik-Jonkman random-effects model
  • Bland-Altman analysis

Pooled result

As reported in the abstract

Agreement of CO monitors vs reference methods

Pooled bias

0.15 L min⁻¹

Limits of agreement (LoA)

± 3.45 L min⁻¹

Pooled percentage error (PE)

49%

Further pooled results

Agreement by device type (subgroup)

Calibrated pulse contour analysis (PCA), PE (best agreement)
25%
Uncalibrated PCA, thoracic electrical bioimpedance and bioreactance, PE (poor agreement)
≥ 52%

Trending ability (acceptable = concordance ≥ 90%)

Datasets reporting trending
15
Datasets achieving concordance ≥ 90%
3

Original publication

Cardiac output monitors in septic shock: do they deliver what matters? A systematic review and meta-analysis. Crit Care. Published 12 July 2025. doi:10.1186/s13054-025-05547-9

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.