Respiratory failure & ventilation

Clinical evaluation of a smart wristband for monitoring oxygen saturation, pulse rate, and respiratory rate

J Clin Monit ComputPublished Diagnostic accuracy study

Authors' conclusion

This study shows that photoplethysmography-based SpO2, RR, and PR measurements can be accurate during heart catheterisations.

Study context

Population
Patients undergoing heart catheterisations945 samples
Participants
100 patients
Comparison
PPG-based wristband vs reference Nellcor fingerclip and 5-lead ECG
Outcome
Accuracy of SpO2, RR and PR vs reference (correlation coefficients and root mean squared error)

Performance

As reported in the abstract

Agreement of wristband vs reference measurements
Test or modelSpO2 rSpO2 root means squared error (RMSE)RR rRR RMSEPulse rate (PR) rPR RMSE
Wristband0.8151.6%0.9760.9 brpm0.9951.3 bpm

Wristband

SpO2 r0.815
SpO2 root means squared error (RMSE)1.6%
RR r0.976
RR RMSE0.9 brpm
Pulse rate (PR) r0.995
PR RMSE1.3 bpm

Further findings

Skin colour and arm hair density subcategories (subgroup)

Similar results across all skin colour and arm hair density subcategories

Original publication

Clinical evaluation of a smart wristband for monitoring oxygen saturation, pulse rate, and respiratory rate. J Clin Monit Comput. Published 10 October 2024. doi:10.1007/s10877-024-01229-z