Pediatric anesthesia & critical care
Comparative efficacy of finger versus forehead Plethysmographic Variability Index monitoring in pediatric surgical patients
Authors' conclusion
Although forehead and finger Plethysmographic Variability Index monitoring were similarly sensitive in predicting fluid responsiveness in pediatric surgical patients, the former method provided higher specificity. The best cut-off value for PVI measurements with forehead and finger probes was found to be 14%.
Study context
- Population
- Mechanically ventilated pediatric patients undergoing minor elective surgery; fluid responsiveness = ≥ 15% increase in CO after passive leg raise
- Participants
- 50 participants
- Comparison
- Finger vs Forehead
- Outcome
- Perfusion index measured at the finger or forehead and PVI monitoring; ability to predict fluid responsiveness
Performance
As reported in the abstract
Prediction of fluid responsiveness (≥ 15% increase in CO after passive leg raise) · prior to passive leg raise maneuver
| Test or model | Value | p | Cut-off | Sensitivity | Specificity |
|---|---|---|---|---|---|
| Finger | AUC 0.699 | p = .011 | |||
| Forehead | AUC 0.847 | p < .001 | |||
| Finger | ≤14% | 92.9% | 45.4% | ||
| Forehead | ≤14% | 96.4% | 72.7% |
Value
FingerAUC 0.699
ForeheadAUC 0.847
Finger— not stated in the abstract
Forehead— not stated in the abstract
p
Fingerp = .011
Foreheadp < .001
Finger— not stated in the abstract
Forehead— not stated in the abstract
Cut-off
Finger— not stated in the abstract
Forehead— not stated in the abstract
Finger≤14%
Forehead≤14%
Sensitivity
Finger— not stated in the abstract
Forehead— not stated in the abstract
Finger92.9%
Forehead96.4%
Specificity
Finger— not stated in the abstract
Forehead— not stated in the abstract
Finger45.4%
Forehead72.7%
Original publication
Comparative efficacy of finger versus forehead Plethysmographic Variability Index monitoring in pediatric surgical patients. Paediatr Anaesth. Published 17 October 2022. doi:10. 1111/ pan. 14561
Composed by AI from the published abstract and checked against it. How Briefs are made →