Pediatric anesthesia & critical care

Determination of reference ranges for the ClotPro® thromboelastometry device in paediatric patients

Br J AnaesthPublished Prospective observational study

Authors' conclusion

ClotPro® reference ranges were determined for six paediatric age groups, and show age-dependent differences in specific parameters.

Study context

Population
Children undergoing elective surgery; venous samples after induction of anaesthesia
Participants
321 children
Exposure
Six paediatric age groups
Outcome
ClotPro® reference ranges (2.5% and 97.5% percentiles) per age group
Methodological detailsreference range determination
  • Reference range determination

Main finding

As reported in the abstract

ClotPro® parameters by age group
EX-test clotting time (CT) significantly shorter at 0-3 months vs all older age groups
P<0.001
EX-test and IN-test A5 and A10 significantly higher in first year of life vs children >1 yr
P<0.001
FIB-test A5 and A10 significantly higher in first 3 months of life
P<0.003

Observational data: associations, not effects.

Original publication

Determination of reference ranges for the ClotPro® thromboelastometry device in paediatric patients. Br J Anaesth. Published 12 November 2022. doi:10.1016/j.bja.2022.09.023