Hemodynamics, fluids & transfusion
DiCARTTM device to measure capillary refill time: a validation study in patients with acute circulatory failure
Authors' conclusion
The DiCART™ device showed poor agreement with clinical CRT in patients with acute circulatory failure
Study context
- Population
- Patients with acute circulatory failure; Capillary Refill Time (CRT) measured on fingertip, chest and knee
- Participants
- 25 patients
- Comparison
- Automated DiCART™ CRT vs clinically measured CRT
- Outcome
- Agreement between CRTDiCART and CRTCLIN (ICC and Bland and Altman analysis)
Methodological detailscomparative interventional study, single-centre, blinded operators, Three measurements per site averaged; clinical and DiCART™ measurements by two different operators, each blinded to the results
- Comparative interventional study
- Single-centre
- Blinded operators
- Three measurements per site averaged; clinical and DiCART™ measurements by two different operators, each blinded to the results
Performance
As reported in the abstract
Agreement CRTCLIN vs CRTDiCART, all measurement sites
Intraclass correlation coefficient (ICC)
0.46
95% CI 0.32–0.59
Agreement CRTCLIN vs CRTDiCART, all measurement sites
Mean bias
0.23 s
95% CI −0.17 to 0.64
upper Limit of Agreement (LoA) 2.77s (95% CI 2.44, 3.20); lower LoA - 2.30s (−2.73, −1.97)
Further findings
ICC
| Outcome | CRTDiCART | CRTCLIN |
|---|---|---|
| Intra-observer agreement (ICC) | 0.43 (95% CI 0.15–0.64) | 0.85 (95% CI 0.74–0.91) |
| Inter-observer agreement (ICC) | 0.41 | 0.86 (95% CI 0.76–0.92) |
- Inter-observer agreement (ICC): CRTDiCART 95% CI lower bound 0.14; upper bound misprinted in abstract
Original publication
DiCARTTM device to measure capillary refill time: a validation study in patients with acute circulatory failure. J Clin Monit Comput. Published 26 February 2025. doi:10. 1007/ s10877-025-01271-5