Cardiac arrest & resuscitation

Does transthoracic impedance influence termination and recurrence of ventricular fibrillation during impedance-compensated shocks for out-of-hospital cardiac arrest? A retrospective study

ResuscitationPublished Retrospective observational study

Authors' conclusion

Despite impedance-compensated defibrillation, TTI ≥100 Ω remains associated with reduced likelihood of VF/pVT termination in OHCA. In contrast, TTI does not appear to influence the risk of recurrent VF/pVT following successful termination.

Study context

Population
Non-traumatic out-of-hospital cardiac arrest (OHCA), patients aged ≥16 years with initial shockable rhythm, January 2019–October 2022, impedance-compensated shocks8,333 analysable shocks
Participants
1,880 participants
Exposure
Transthoracic impedance <100 Ω vs ≥100 Ω, per 1-Ω increase
Outcome
Termination of VF/pVT (no shockable rhythm within 5 s after shock, resulting in asystole or organised rhythm)
Methodological detailsmixed-effects logistic regression, shock-level analysis
  • Mixed-effects logistic regression
  • Shock-level analysis

Main finding

As reported in the abstract

VF/pVT termination
Adjusted odds ratio per 1-Ω increase (95% CI)P value
TTI <100 Ω
0.999(0.993–1.006)
P = 0.860
TTI ≥100 Ω
0.994(0.988–0.999)
P = 0.016

Observational data: associations, not effects.

Further findings

Recurrent VF/pVT after termination

aOR per 1-Ω increase (95% CI)P value
TTI1.003 (0.999–1.006)P = 0.082

Original publication

Does transthoracic impedance influence termination and recurrence of ventricular fibrillation during impedance-compensated shocks for out-of-hospital cardiac arrest? A retrospective study. Resuscitation. Published 25 May 2026. doi:10.1016/j.resuscitation.2026.111151

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.