Respiratory failure & ventilation

Dynamic mechanical power at the bedside: A validation study in volume-controlled and pressure-controlled ventilation modes

J Crit CarePublished Prospective observational validation study

Authors' conclusion

MPdyn demonstrates high agreement with MPgeo in both VCV and PCV modes, supporting its use as a simple, reliable, and bedside-applicable tool for calculating mechanical power in ARDS patients.

Study context

Population
Adults with ARDS in a tertiary ICU, measured in VCV and pressure-controlled ventilation (PCV) at I:E 1:2 and 1:1
Participants
36 adults
Comparison
Dynamic mechanical power (MP) equation vs geometric gold standard
Outcome
Agreement of MPdyn with MPgeo (linear regression, Bland-Altman bias)
Methodological detailssingle tertiary ICU, Bland-Altman analysis
  • Single tertiary ICU
  • Bland-Altman analysis

Performance

As reported in the abstract

Agreement MPdyn vs MPgeo (bias MPdyn - MPgeo)
Correlation, all modes and I:E ratios (R2)
≥ 0.97
Bias, VCV 1:2
0.15 J/min (0.8 %)
Bias, VCV 1:1
0.32 J/min (2.1 %)
Bias, PCV 1:2
0.28 J/min (1.9 %)
Bias, PCV 1:1
0.52 J/min (3.4 %)
SD of bias across analyses
<0.5 J/min

Original publication

Dynamic mechanical power at the bedside: A validation study in volume-controlled and pressure-controlled ventilation modes. J Crit Care. Published 30 October 2025. doi:10.1016/j.jcrc.2025.155314

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.