Pediatric anesthesia & critical care

Evaluation of the Noninvasive Estimated Continuous Cardiac Output System for Pediatric Patients: A Prospective Observational Study

Anesth AnalgPublished Prospective observational study

Authors' conclusion

Cardiac output measurement by esCCO with modified coefficients for pediatric patients showed high accuracy and tracking ability compared with cardiac output measurement by transthoracic echocardiography.

Study context

Population
Pediatric patients aged <15 years undergoing elective noncardiac surgerymain study
Participants
80 patients
Exposure
Modified estimated continuous cardiac output (esCCO) vs transthoracic echocardiography for cardiac output
Outcome
Accuracy and trending ability of cardiac output measurement (Bland-Altman analysis and polar plot)
Follow-up
Pre- and postoperatively
Methodological detailsmethod-comparison, Bland-Altman analysis, polar plot
  • Method-comparison
  • Bland-Altman analysis
  • Polar plot

Main finding

As reported in the abstract

Agreement of modified esCCO with transthoracic echocardiography
Correlation coefficient, preoperative
0.96
Correlation coefficient, postoperative
0.98
Bias (SD), preoperative
0.03 (0.28)
95% limits of agreement, preoperative
−0.53 to 0.60
Percentage error, preoperative
18%
Bias (SD), postoperative
−0.04 (0.19)
95% limits of agreement, postoperative
−0.42 to 0.35
Percentage error, postoperative
15%
Polar plot angular bias (SD)
2.9° (6.0°)
Polar plot radial 95% limits of agreement
−9.2° to 14.9°

Observational data: associations, not effects.

Original publication

Evaluation of the Noninvasive Estimated Continuous Cardiac Output System for Pediatric Patients: A Prospective Observational Study. Anesth Analg. Published 4 September 2024. doi:10.1213/ane.0000000000007144

Composed by AI from the published abstract and checked against it. How Briefs are made →