Pediatric anesthesia & critical care

Objective real-time epidural pressure measurement using the CompuFlo® device, a mono-center observational study

Anaesth Crit Care Pain MedPublished Prospective observational study

Authors' conclusion

The CompuFlo® device allowed a pressure measurement prior to obtaining LOR during EA catheter placement in children. Younger children appear to have lower pressure levels.

Study context

Population
Children requiring epidural anesthesia for surgery
Participants
47 children · 56 procedures
Outcome
Pressure levels just before loss of resistance during epidural anesthesia (CompuFlo® device)
Methodological detailsmono-center
  • Mono-center

Main finding

As reported in the abstract

Last pressure just before LOR
Median pressure before loss of resistance (LOR) (quartiles), mmHg
174 (120; 314)
Correlation of pressure before LOR with weight (automatic group)
r = 0.61
Pressure significantly lower in children under 5 kg vs other weight groups
p = 0.02

Observational data: associations, not effects.

Further findings

Correlation of pressure before LOR with weight

Correlation with weight (automatic group), r
0.61
Pressure significantly lower in children under 5 kg vs other weight groups
P = 0.02

Original publication

Objective real-time epidural pressure measurement using the CompuFlo® device, a mono-center observational study. Anaesth Crit Care Pain Med. Published 1 May 2025. doi:10.1016/j.accpm.2025.101530

Composed by AI from the published abstract and checked against it. How Briefs are made →