Objective Real‐Time Peripheral Regional Anesthesia Pressure Measurement in Children Using the CompuFlo Device, a Mono‐Center Observational Study
Authors' conclusion
Pressure does not appear to be the main factor prompting the operator to continue or stop the injection.
Study context
- Population
- Children scheduled for peripheral nerve block175 injections analysed
- Participants
- 47 children
- Outcome
- Maximal injection pressure for each injection during PNB (CompuFlo device)
Methodological detailsblinded, single-centre
- Blinded
- Single-centre
Main finding
As reported in the abstract
Maximum injection pressure
Mean maximum injection pressure, mmHg
515
95% CI 478–551
Injections exceeding 750 mmHg threshold
16%
Observational data: associations, not effects.
Further findings
Factors influencing injection pressure
- Operator experience, needle diameter, peripheral nerve block (PNBs) in restricted diffusion space or lack of injection precision, and incorrect ultrasound injection positions
Original publication
Objective Real‐Time Peripheral Regional Anesthesia Pressure Measurement in Children Using the CompuFlo Device, a Mono‐Center Observational Study. Paediatr Anaesth. Published 14 January 2026. doi:10. 1002/ pan. 70113