Paired assessment of two electromyographic neuromuscular monitors: stimpod NMX450X versus Datex-Ohmeda E-NMT

J Clin Monit ComputPublished Method comparison study

Authors' conclusion

In routine clinical use, the Stimpod NMS450X and Datex-Ohmeda E-NMT demonstrated poor concordance and high variability, likelyinfluenced by device-specific artifacts, connectivity limitations and electrode sensitivity.

Study context

Population
Adults undergoing elective non-cardiac surgery
Participants
24 adults
Comparison
Stimpod NMS450X vs Datex-Ohmeda E-NMT electromyography (EMG) neuromuscular monitors
Outcome
Agreement between devices (Bland-Altman analysis, repeatability coefficients, concordance correlation coefficient) for TOF counts and TOF ratios recorded at 1-minute intervals
Methodological detailspaired, contralateral simultaneous monitoring
  • Paired
  • Contralateral simultaneous monitoring

Main finding

As reported in the abstract

Inter-device agreement of TOF count and TOF ratio
GroupAverage train-of-four (TOF)-count values per patientAverage TOF-ratio values per patient
Datex-Ohmeda12171
Stimpod5639

Average train-of-four (TOF)-count values per patient

Datex-Ohmeda121
Stimpod56

Average TOF-ratio values per patient

Datex-Ohmeda71
Stimpod39
Repeatability coefficient, 95% CI 0.80-0.83 (poor precision)
0.81
Concordance correlation coefficient (CCC) (weak agreement)
0.662

Original publication

Paired assessment of two electromyographic neuromuscular monitors: stimpod NMX450X versus Datex-Ohmeda E-NMT. J Clin Monit Comput. Published 3 February 2026. doi:10.1007/s10877-026-01411-5

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.