Anesthetic drugs & techniques

Precision of electromyography according to the calibration approach of neuromuscular monitoring: a randomised prospective agreement study

J Clin Monit ComputPublished Randomised prospective agreement study

Authors' conclusion

Electromyography was most precise when calibration was performed under deep anaesthesia (control).

Study context

Population
Patients undergoing anaesthesia with quantitative neuromuscular monitoring by electromyography
Participants
60 patients
Comparison
Before induction vs During induction vs Uncalibrated vs Control
Primary endpoint
Repeatability coefficient (fluctuation of the following train-of-four (TOF) reading; precision)
Follow-up
TOF ratios ≥ 0.8
Methodological detailsmulticentre
  • Multicentre

Primary result

As reported in the abstract

Repeatability coefficient at TOF ratios ≥ 0.8

Uncalibrated

0.124

± 0.130

During induction

0.087

± 0.104

Before induction

0.075

± 0.036

Control (contralateral arm) in the uncalibrated, during-induction and before-induction patients, respectively
0.072 ± 0.027, 0.061 ± 0.021, and 0.083 ± 0.063

Secondary outcomes

Recall of pain at anaesthesia induction

Did not differ between investigational groups

Original publication

Precision of electromyography according to the calibration approach of neuromuscular monitoring: a randomised prospective agreement study. J Clin Monit Comput. Published 28 May 2025. doi:10.1007/s10877-025-01304-z

Composed by AI from the published abstract and checked against it. How Briefs are made →