Hemodynamics, fluids & transfusion

Reliability of Noninvasive Blood Pressure Monitoring During Atrial Fibrillation Ablation Under General Anesthesia

J Cardiothorac Vasc AnesthPublished Retrospective observational cohort study

Authors' conclusion

While rhythm did not significantly alter NIBP reliability, substantial variability-particularly in systolic measurements-limits its interchangeability with IBP during AF ablation.

Study context

Population
Adults undergoing endovascular atrial fibrillation (AF) ablation under GA, tertiary academic medical center, June 2022–March 20231,158 paired measurements
Participants
90 adults
Exposure
Noninvasive vs invasive BP monitoring
Outcome
Bias of NIBP vs IBP
Methodological detailssingle-centre
  • Single-centre

Main finding

As reported in the abstract

Noninvasive blood pressure (NIBP) underestimated SBP (all p < 0.05)
3.5 mmHg
NIBP overestimated diastolic BP (all p < 0.05)
2.3 mmHg
NIBP overestimated MAP (all p < 0.05)
1.4 mmHg
Within-subject variability highest for SBP (SD)
12.3 mmHg
Intraclass correlation coefficients across BP types
0.73 to 0.84

Observational data: associations, not effects.

Further findings

Effect of cardiac rhythm on agreement

Cardiac rhythm did not significantly affect agreement; AF and paced rhythm associated with increased systolic error in univariate models

Other influences on discrepancies

Ventilation mode and vasopressor use also influenced discrepancies

Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.

Original publication

Reliability of Noninvasive Blood Pressure Monitoring During Atrial Fibrillation Ablation Under General Anesthesia. J Cardiothorac Vasc Anesth. Published 6 March 2026. doi:10.1053/j.jvca.2026.03.003

Composed by AI from the published abstract and checked against it. How Briefs are made →