Respiratory rate measurement by pressure variation in the high flow nasal cannula-system in healthy volunteers

J Clin Monit ComputPublished Method comparison study

Authors' conclusion

The pressure variation in the HFNC system allows for respiratory rate and effort monitoring, but requires further development to increase precision.

Study context

Population
Healthy volunteers breathing regularly, quickly and slowly on modified HFNC system
Participants
25 volunteers
Comparison
Pressure-based respiratory rate count in HFNC system vs count by observation
Outcome
Agreement of respiratory rate counted from pressure variation vs observation

Main finding

As reported in the abstract

Agreement of pressure-based vs observed respiratory rate (95% Limits of Agreement)
Group95% Limits of Agreement
30 l/min -closed mouth- 1.85;1.91
70 l/min - closed mouth−2.25;2.47
70 l/min - open mouth−30.32;19.93

95% Limits of Agreement

30 l/min -closed mouth- 1.85;1.91
70 l/min - closed mouth−2.25;2.47
70 l/min - open mouth−30.32;19.93

Further findings

Outcome30 l/min - open mouth70 l/min - open mouth
False detection of apnea in pressure-based measurements10%11%

False detection of apnea in pressure-based measurements

Not with a closed mouth

Correlation of pressure difference with respiratory effort

Correlation present, except for 30 l/min with open mouth

Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.

Original publication

Respiratory rate measurement by pressure variation in the high flow nasal cannula-system in healthy volunteers. J Clin Monit Comput. Published 12 June 2024. doi:10.1007/s10877-024-01185-8

Composed by AI from the published abstract and checked against it. How Briefs are made →