Authors' conclusion

The temperature recorded by iThermonitor has better concordance with the core temperature than the bladder temperature recorded by Foley catheter sensor in laparoscopic rectal surgery.

Study context

Population
Adults undergoing laparoscopic rectal surgery
Participants
82 adults · 3,303 measurement pairs
Comparison
iThermonitor vs Bladder vs Oesophageal
Outcome
Concordance with core (oesophageal) temperature: Bland-Altman mean difference and limits of agreement
Follow-up
Whole surgical period
Methodological detailsrepeated-measures Bland-Altman analysis
  • Repeated-measures Bland-Altman analysis

Main finding

As reported in the abstract

Agreement with oesophageal temperature

Mean difference

0.05 °C

iThermonitor vs Oesophageal · limits of agreement - 0.48 to 0.56 °C

Mean difference

0.28 °C

P < 0.001

Oesophageal vs Bladder · limits of agreement - 0.39 to 0.94 °C; P from F-test vs. iThermonitor

Further findings

Agreement during pneumoperitoneum period

Similar results during the pneumoperitoneum period

Original publication

Temperature measurements of a wearable and wireless axillary sensor iThermonitor but not a bladder probe represents the core temperature during laparoscopic rectal surgery. J Clin Monit Comput. Published 5 July 2022. doi:10.1007/s10877-022-00892-4

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.