Temperature measurements of a wearable and wireless axillary sensor iThermonitor but not a bladder probe represents the core temperature during laparoscopic rectal surgery
Authors' conclusion
The temperature recorded by iThermonitor has better concordance with the core temperature than the bladder temperature recorded by Foley catheter sensor in laparoscopic rectal surgery.
Study context
- Population
- Adults undergoing laparoscopic rectal surgery
- Participants
- 82 adults · 3,303 measurement pairs
- Comparison
- iThermonitor vs Bladder vs Oesophageal
- Outcome
- Concordance with core (oesophageal) temperature: Bland-Altman mean difference and limits of agreement
- Follow-up
- Whole surgical period
Methodological detailsrepeated-measures Bland-Altman analysis
- Repeated-measures Bland-Altman analysis
Main finding
As reported in the abstract
Agreement with oesophageal temperature
Mean difference
0.05 °C
iThermonitor vs Oesophageal · limits of agreement - 0.48 to 0.56 °C
Mean difference
0.28 °C
P < 0.001
Oesophageal vs Bladder · limits of agreement - 0.39 to 0.94 °C; P from F-test vs. iThermonitor
Further findings
Agreement during pneumoperitoneum period
- Similar results during the pneumoperitoneum period
Original publication
Temperature measurements of a wearable and wireless axillary sensor iThermonitor but not a bladder probe represents the core temperature during laparoscopic rectal surgery. J Clin Monit Comput. Published 5 July 2022. doi:10. 1007/ s10877-022-00892-4