Cardiothoracic & vascular anesthesia
The Use of Contrast May Improve Aortic Valve Assessment During Transesophageal Echocardiography
Authors' conclusion
This discrepancy is significant and could affect the decision to replace the aortic valve.
Study context
- Population
- Patients undergoing surgical aortic valve replacement for a stenotic valve, tertiary hospital
- Participants
- 30 consecutive patients
- Exposure
- Contrast vs noncontrast transesophageal echocardiography (TEE) Doppler assessment of the aortic valve
- Outcome
- Aortic valve area
Methodological detailssingle-centre, retrospective review, consecutive patients, Bland-Altman analysis
- Single-centre
- Retrospective review
- Consecutive patients
- Bland-Altman analysis
Main finding
As reported in the abstract
Without contrast agent, aortic valve area overestimated vs TTE
0.26 cm²
Observational data: associations, not effects.
Further findings
Peak and mean aortic valve gradients without contrast
- Peak gradient underestimated (mmHg)
- 19
- Mean gradient underestimated (mmHg)
- 11
- p value <0.001
Pulse-wave Doppler V1 velocity of the left ventricular outflow tract (LVOT)
- Not affected by contrast
Parts of this Brief are withheld: the automatic checks could not confirm them against the abstract.
Original publication
The Use of Contrast May Improve Aortic Valve Assessment During Transesophageal Echocardiography. J Cardiothorac Vasc Anesth. Published 22 February 2023. doi:10. 1053/ j. jvca. 2023. 02. 023
Composed by AI from the published abstract and checked against it. How Briefs are made →