A novel adaptive filter with a heart-rate-based reference signal for esophageal pressure signal denoising
Authors' conclusion
The proposed technique can effectively eliminate the CGOs from the measured Peso signals in real-time without requiring additional equipment to collect the reference signal.
Study context
- Population
- ICU patients receiving mechanical ventilation8 on controlled ventilation analysed
- Participants
- 20 patients
- Comparison
- Adaptive filter with heart-rate-based sine reference signal vs unfiltered esophageal pressure
- Outcome
- CGO removal from Peso
Methodological detailsreal-time adaptive filtering, clinical data
- Real-time adaptive filtering
- Clinical data
Main finding
As reported in the abstract
Reduction in SD/mean of lung compliance, without changing compliance (8 patients, controlled ventilation)
28.4-79.2%
Original publication
A novel adaptive filter with a heart-rate-based reference signal for esophageal pressure signal denoising. J Clin Monit Comput. Published 4 February 2024. doi:10. 1007/ s10877-023-01116-z