A novel adaptive filter with a heart-rate-based reference signal for esophageal pressure signal denoising

J Clin Monit ComputPublished Methodological study

Authors' conclusion

The proposed technique can effectively eliminate the CGOs from the measured Peso signals in real-time without requiring additional equipment to collect the reference signal.

Study context

Population
ICU patients receiving mechanical ventilation8 on controlled ventilation analysed
Participants
20 patients
Comparison
Adaptive filter with heart-rate-based sine reference signal vs unfiltered esophageal pressure
Outcome
CGO removal from Peso
Methodological detailsreal-time adaptive filtering, clinical data
  • Real-time adaptive filtering
  • Clinical data

Main finding

As reported in the abstract

Reduction in SD/mean of lung compliance, without changing compliance (8 patients, controlled ventilation)

28.4-79.2%

Original publication

A novel adaptive filter with a heart-rate-based reference signal for esophageal pressure signal denoising. J Clin Monit Comput. Published 4 February 2024. doi:10.1007/s10877-023-01116-z