Comparison of mainstream end tidal carbon dioxide on Y-piece side versus patient side of heat and moisture exchanger filters in critically ill adult patients: a prospective observational study

J Clin Monit ComputPublished Prospective observational method comparison study

Main finding · Accuracy difference (Y-piece vs patient side)Result — go to the resultAccuracy difference - 0.14 mmHg two-sided 90% CI - 0.58 to 0.29

Authors' conclusion

The accuracy of mainstream EtCO2 measurements on the Y-piece side of HMEF was noninferior to that on the patient side in critically ill adults.

Study context

Population
Critically ill adults receiving mechanical ventilation with heat and moisture exchanger filter (HMEF)
Participants
37 adults
Exposure
Mainstream EtCO2 on Y-piece side vs patient side of HMEF
Outcome
Accuracy of EtCO2 (absolute difference between PaCO2 and EtCO2); non-inferiority margin + 1 mmHg
Methodological detailsnon-inferiority, July 2019 to December 2019
  • Non-inferiority
  • July 2019 to December 2019

Main finding

Accuracy difference (Y-piece vs patient side)

Accuracy difference

- 0.14 mmHg

two-sided 90% CI - 0.58 to 0.29

P < 0.001 for non-inferiority

As reported in the abstract

Observational data: associations, not effects.

Further findings

Agreement between PaCO2 and EtCO2 (Bland-Altman 95% limits of agreement)

Limits of agreement similar on both sides of HMEF

Original publication

Comparison of mainstream end tidal carbon dioxide on Y-piece side versus patient side of heat and moisture exchanger filters in critically ill adult patients: a prospective observational study. J Clin Monit Comput. Published 3 August 2022. doi:10.1007/s10877-022-00901-6

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.